Features
1.Windows operation interface, figure button, mouse operation
2.96-well plate, multiple tests in the single plate
3. Arrangement of Test item in Micro-plate can be store and re-use
4. Stability: <0.005Abs/10min
5. Linearity: r > 0.999
6.Repeatability: < 0.2%
7. Plate shaking, speed and time adjustable
8.Open system, test item and parameters can be added or modified
9. Large storage for 1000 test item and 1,000,000 results
10.External large LCD color screen
11. Printing comprehensive report including hospital and patient information and test results
12.External computer, easy to use
Technical Specifications
Measuring system |
8-channel optical system |
Test mode |
End-point, kinetic, Multi-wave analysis |
Micro-plate types |
96-well plate, multiple tests in the single plate |
Wavelength |
405nm,450nm,492nm,630nm,4 free positions for additional filters |
Wavelength precision |
2 nm, width ≤ 10nm |
Light source |
Tungsten Halogen,12V/20W,>2000 hours |
Reading range |
0~4.000 Abs |
Measurement range |
0~3.000 Abs |
Resolution |
<1% or 0.001Abs |
Stability |
<0.005Abs/10min |
Well Coincidence |
≤0.015A |
Linearity |
r > 0.999 |
Repeatability |
< 0.3% |
Plate shaking function |
Plate shaking, speed and time adjustable |
Quality control curve |
Westguard, Multi-rule, automated QC alert |
Reagents |
Open system
test item and parameters can be added or modified |
Storage |
1000 test item and 1,000,000 results |
Printer |
External jet printer, multi-form patient report |
Display |
10 inch color LCD |
Language |
English |
Power supply |
220VAC 15% or 110VAC 15%50-60Hz 100w |
Machine dimension |
43cm(L)X34cm(W)X16cm(H) |
Work environment |
temperature15°C-30°C,Wet≤90% |
Weight |
9Kg |