Technical Specifications
1. General
Monitoring Parameters: SpO2、PR、PLETH、Pulse strength
Signal sockets:SpO2 socket、PR socket、USB socket(for AC power)
2. Display screen
Type: 240 × 320 dot matrix color TFT LCD screen
Display area: 43mm×58mm
3. Functions
PLETH: The oximeter can display PLETH waveform
Audio and visual alarm: The speaker sounds the alarm and the parameter’s measurement will become in red to flash on the screen. Also the same when sensor-off.
Prompting: Icons of battery capacity, pulse sound switch and alarm sound switch.
Power saving: The oximeter will automatic shut down to save power. The time is settable.
4. Power Requirements
AC power adaptor
Input: AC 100 V~240V, 50/60Hz, 5VA
Output: D.C. 5V
Internal battery
Type: D.C. 3.7V/3000mAH Li-ion rechargeable battery
Run Time: 15-hour continuous operation with a new, fully charged battery (environment temperature is 25℃)
Recharge time: 8 hours
5. Environment
Temperature
Operation: 0°C ~ 45 °C
Transportation and storage: -20°C ~ 60 °C
Humidity
Operation: 15% - 95 %( noncondensing)
Transportation and storage: 10% - 95 % (noncondensing)
Altitude
Operation: 86 KPa ~106 KPa
Transportation and storage: 50KPa~106 KPa
6. Parameter Specifications
SpO2
Patient: Adult, pediatric,neonate
Range: 0% - 100%
Resolution: 1 %
Accuracy: 70% - 99%: ±2 %
0% - 69%: Unspecified
PR
Range: 0 bpm ~250bpm
Resolution: 1bpm
Accuracy: 2bpm