Features
1.7 inch Color LCD , Touch screen
2.Flow cell and Cuvette test mode compatible
3.Real time curve showing
4.7 wavelength , 90 test item pre-programmed
5.Memory for 10,000 sample results
6.Light source hibernate function making lamp life longer
7.RS232 interface, PC connecting
Test contents
Liver Function:
Alanine Aminotransferase(ALT), Glutamic-oxalacetic Transaminase(AST), Alkalinity Phosphatase(ALP), Total Bilirubin(TBIL),
Direct Bilirubin(DBIL), Total Protein(TP), Albumin(ALB), TTT,CHE,Nh
Renal Function:
BUN, Inosine Anhydride(CRE), Urea Nitrogen(URE),UA,CO2-CP, Y-GT
Lipids:
Cholesterol(TCHO), Triglyceride(TG),HDL-C,LDL-C
Cardiac Function:
CK,CK-MB,LDH
Immune Proteins:
IGA,IGG,IGM,C3
Other Function:
Glucose(GLU),HGB,AMS
Specifications
Test mode: Kinetic, End point, Two point, Absorbance
Wavelength: 7 standard filters, 340,405,492,546,578,630 nm,1 free position
Wavelength precision: ±2 nm, width ≤ 10nm
Absorbance range: 0~4.500 Abs
Absorbance precision: outside 0.0001 Abs , inside 0.00001 Abs
Carry over: < 1%
Repeatability CV: < 1%
Incubator: 37℃,30℃,25℃
Precision : ±0.1℃
Sample volume: 0~3000 uL (≥500uL recommend)
Data Storage: ≥10000
Quality control curve: SD,CV%,L-J quality control curve for every test item
Flow cell: 32 uL quartz glass, 10 mm
Lamp: Philips 6V 10W Halogen Lamp with long life
Screen: 7 inch color LCD
CPU: ARM series high speed micro CPU
Printer: Internal thermal sensitive printer
Interface: RS-232
Work Environment: Temperature 0℃ ~ 40℃, Humidity: ≤80%
Power supply: 100~240VAC,50-60Hz,100VA
Dimension: 340(L)*270(W)*160(H)
Weight: 5 kg