 ECG    
Lead mode: 3 Leads or 5 Leads 
Lead selection: I,II,III,AVR,AVL,AVF,V 
Gain:X0.25,X0.5,X1,X2,Auto 
Heart Rate:15-300BPM(Adult) 15-350BPM(Neonatal) 
Resolution:1 BPM 
Accuracy:±1% 
ST measurement Range: -2.0+2.0mV 
Pacemaker: Yes 
ARR detecting: Yes 
Electrode offset potential: ±300mV 
Leakage current: <10uA 
  
NIBP 
Work mode: Manual, Auto, Continuous 
Measure Range: 
Adult:  SYS:  40-270mmHg 
          DIA:  10-215mmHg 
          MEAN:20-235mmHg 
Pediatric: 
        SYS:40-200mmHg 
        DIA: 10-150mmHg 
        MEAN:20-165mmHg 
Neonatal:  
        SYS:40-135mmHg 
        DIA:10-100mmHg 
        MEAN:20-110mmHg 
Resolution:1mmHg 
Measurement Type :SYS,DIA,MEAN  | 
 SpO2 
Measure Range:0-100% 
Resolution:1% 
Accuracy:2%(70-100%),,0-69%,upspecified 
Pulse Rate:20-300BPM 
Resolution:1BPM 
Accuracy:±2BPM 
  
RESP 
Measure Range:0-150RPM 
Resolution:1BrPM 
Accuracy:±2% 
  
TEMP*2 
Measure Range:0-50C 
Resolution:0.1C 
Accuracy:±0.1C 
  
IBP: 
Measure range: 
ART:0-300 mmHg 
PA:-6-120 mmHg 
CVP/RAP/LAP/ICP:-10-40 mmHg 
P1/P2:-10-300 mmHg 
Accuracy: ±2% or ±1 mmHg 
  
Environment: 
Operating Temperature: 59°to 95°F(15°to 35°C) 
Storage Temperature: 23°to 122°F(-5°to 50°C) 
Operating Humidity: 15℅ to 90℅,non-condensing  |